PLS-100偏振光源检测仪器和PEM-330(偏振消光比计)组合提供最JIA的面向偏振相关光器件产品的研发及产品线测试的解决方案,偏振消光比测试范围大于40dB。
- 偏振消光比可以达到45dB(典型值)
- 高输出光功率、稳定性高
- 波长范围1310±20nm或者1550±20nm
- 光学元器件的偏振周调整
- 保偏光纤接头固定作业、测试等
Parameter
Unit
Specification
Notes
Wavelength Range nm
1310±20 or 1550±20
Polarization Extinction Ratio (PER) dB
40
45dB typically
PER Stability @1hour *1 dB
±0.5
±0.1 typically
Output Power dBm
0
+3dBm typically
Power Stability @1hour *1 dB
<±0.1
±0.05 typically
Power Variation by changing angle dB
±0.2
Linewidth (FWHM) *2 nm
>20
Optical Connector –
FC or SC
Connector Polish –
SPC
Polarized Axis Adjustment deg
2
Fine adjustment
deg
180
Coarse adjustment
Operating Temperature ℃
15 to 35
Size (W) x (D) x (H) mm
210x350x135
except for the protruding portion
Weight kg
3.5
Supply Voltage V
AC100-240 ±10%
Frequency Hz
50/60
*All measurements are taken after 60 min warm up and at stable room temperature.
*1: For period of 1 hour. Within ± 0.5 ºC.
*2: Measured at Full Width Half MaximumPLS-100偏振光源检测仪器
所有评论仅代表网友意见,与本站立场无关。